SAFETY INSTRUCTIONS
6
•
Accessibility of the power adaptor.
The plug
socket must be close to the power adaptor
and be easily accessible.
•
Internal bleeding.
Proceed with caution if
you are prone to internal bleeding; for
example, after an injury or a fracture
WARNING
The Revolution Wireless must only be operated
in dry rooms.
•
Do not use
the Revolution Wireless in water
or in a humid atmosphere (sauna, Bath,
Shower etc.) that would cause electronic
failure.
•
Water protection.
The unit is not protected
from the ingress of water
•
When connecting the unit to other
equipment or when creating a medical
system, check that the sum of leakage
currents will not cause any hazard. Please
contact DJO GLOBAL if you have questions
regarding this matter.
•
No modification of this equipment is
allowed.
•
Do not open the product and its accessories
as there is risk of electrocution
•
Before cleaning and service interventions,
turn the remote control and the modules off
and disconnect the tablet from the power
line by removing the power cord from the
wall outlet.
•
Liquids and foreign material (such as dust,
metal etc.) must not be allowed
to enter
the remote control, the modules, the
charging tablet or the power supply. If such
material has entered into the units, it must
be immediately checked by a service
technician, before it can be reused.
•
Electricity supply.
Never connect the
stimulation cables to an external power
supply, as there is a risk of electrocution.
•
Do not apply stimulation near the area of an
implant, such as cochlear implants,
pacemakers, skeletal anchorage or electric
implants. This could cause an electrical
shock, burns, electrical interference or
death.
•
Never use the Revolution Wireless or the AC
adaptor if it is damaged or open. There is a
risk of electric shock.
•
Disconnect the AC adaptor immediately if
there is abnormal heating or smell, or if
smoke comes from the AC adaptor or the
device.
•
Do not place the docking station in a
confined space (carrying case, drawer etc.)
when charging the device. There is a risk of
electrocution.
WARNING
Equipment malfunction —
these warnings refer
to potential equipment mailfunctions that could
result in patient hazards
•
Magnetic and electrical fields are capable of
interfering with the proper performance of
the unit. For this reason make sure that all
external devices operated in the vicinity of
the unit comply with the relevant EMC
requirements. ¬ Devices that intentionally
radiate electromagnetic energy, such as X-
ray equipment, MRI devices, ¬radio systems
and cell phones are possible sources of
interference, as well as RFID equipment (e.g.
electromagnetic security systems).Keep the
unit away from such equipment and verify
its performance before use. Foradditional
information, refer to the EMC section of this
Manual.
•
Do not use the Revolution Wireless within 3
feet of short wave or microwave devices as
this could alter the currents generated by
the stimulator. If you are in any doubt as to
the use of the stimulator in close proximity
to another medical device, seek advice from
the manufacturer of the latter or from your
doctor.
WARNING
•
Exercise caution when using electrotherapy
while the patient is connected to monitoring
equipment with electrodes attached to the
body. Stimulation could disrupt the signals
sent to the monitoring equipment.