DEVICE ISSUES
31
The module’s battery has run out
The remote control shows a discharged battery symbol. Shown on the channel
where the problem was detected (in this case channel 4).
The stimulation is not producing the usual sensation
The stimulation causes discomfort
Problem
Possible cause
Solution
Low module
battery
During the
stimulation a module
may turn out to be
discharged.
1.
You can stop the stimulation and recharge the module with the empty
battery.
2.
You can set aside the module with the empty battery and continue your
stimulation session without it.
Problem
Possible cause
Solution
Unusual
stimulation
sensation
Improper electrode
placement
Low intensity
1.
Check that all the settings are correct and ensure that the electrodes are
positioned properly.
2.
Change slightly the positioning of the electrodes. Electrodes must be a
minimum of 2 inches apart.
3.
Replace the electrodes
4.
If the output is too weak, check that the channels are active, and then
increase intensity
Problem
Possible cause
Solution
Stimulation is
uncomfortable
Amplitude (intensity)
is too high or wrong
program
Electrodes are too
close together
Damaged or worn
electrodes or lead
1.
Check that all the settings are correct and ensure that the electrodes are
positioned properly.
2.
The electrodes are losing their bonding strength and are no longer
providing suitable contact with the skin.
3.
Reposition the electrodes. Electrodes must be a minimum of 2 inches
apart.
Note: Final accessories may differ from the picture