VERIFICATION PROCEDURE
44
3.
From the home menu, enter into the “Settings”. Then
enter into the “Pair a new module” option. According
to the lack of the green circles on the right top of the
screen, no module should be detected for the
moment.
4.
In order to validate the wireless communication, at
this stage, turn on all four functional modules and
pair them with the remote.
For each module, verify that it can be paired
(the On/Off button LED briefly switches from
green to red when it’s turned on before being
permanently green).
For each module, verify that the synchronization
takes less than 4 seconds.
When all the modules are paired, verify that all the
empty circles at the top of the remote screen are now
green.
5.
From the home menu, choose the “TENS 100Hz 1”
program (the same as before), select and enter into
the stimulation program.
Make sure that every module is turned on and
verify that every module is linked to the remote.
6.
Take the body load device and place a module on it.
Switch on this module. Turn off the three other
modules.
Launch the session by pressing the “start” option and by
increasing the intensity of the module on the body load.
Once that the stimulation session has begun,
verify that the On/Off button of the module
blinks orange.
Note: Final accessories may differ from the picture
Note: Final accessories may differ from the picture
Note: Final accessories may differ from the picture