ERROR CODES
38
Error are defined as being abnormal parameter or behavior of the module (including both hardware and software issues)
which firmware can’t recover from in a safe and functional way. The increase of error will stop all safety critical process, and
store a 16bits error code will flag the abnormal parameter or behavior of the module (including both hardware and
software issues). The initialization self-tests will also generate an error code. If the paired RC is in range, the module will
communicate the error code.
Code
Description / Comment
Part to change
0x4007
C_ErrorCpxRF01BadFwInfoChunk. RF reset error
Remote
0x271F
ERROR_PROGMGR_WRONGOPMODE
0x00CF
This might be an out of range side-effect
0x4104
RFNode not properly reset. Check (triple) the RF reset from remote.
Remote
0x1140
C_ErrorDevlibProgram
0x410A
SPI CRC error Menu 0x2900 (Menu poweroff) once, 0x6100 (menu charge). RF not (yet) reset when in
menu ext. power on. Mask RF errors on Menu_ExtPowerOn
Remote
0x1180
C_ErrorDevlibNoStimBarImg
Remote FW
0x410D
C_ErrorCpxRF01DMATEError
0x011B
ERROR_BUTTON_NEVER_PRESSED
Module
0x00DC
Malformed pulse, signal IPH/INH not conform. Check what's happening when stop/emerstop is received
exactly during pulse
Module (80/20
FW vs HW)
0x00D4
HV too high (>195V @ anytime during module use)
0x4904
When remote is preparing a communication to 1 or several modules, the target is not defined (none
selected). : 1. The menu in which the command was called missed a check on number of modules. Menu
is 0xf0203, C_MenuId_RfTest, only accessible from menu_rd !!! Check if this error could have been raised
before buy. If not, hide the Menu RD
0x4102
C_ErrorCpxRF01HwFault
0x0200
Bug with "pausing mi-scan". miscan not deinitialized when pausing it
Module FW
0x4753
side effects of 0x480B
0x4737
side effects of 0x480B
0x3001
C_ErrorRecordVerify
Remote
0x271C
Jump sent to a wrong program ID