VERIFICATION PROCEDURE
43
REMOTE CONTROL AND MODULE
INTERACTION
The purpose of this verification procedure is to check the
functionalities of the replaced remote control and the
communication with modules.
Tools and equipment
•
Remote control
•
4 modules
•
Body load
Procedure
1.
Switch on the remote control with the On/Off button.
Verify that the remote displays the device
welcome screen and then the daily task screen
or the select language screen (if the remote has
the initial version).
2.
Once the remote control is switched on, a
combination of the user interface buttons need to be
done in order to test the functionality of the user
commands. The following procedure contains the
combination of the buttons and their associated
actions.
Verify the correct response of the multi-function
buttons.
Verify the action of each multi-directional
button.
Verify the functionality of the validation button.
Verify the illumination of every guide light when
their related button can allow a user interface
action.
Starting from the home menu > “Favorites”
Pressed button
Button action
3x Down direction
“Rehabilitation I&II”
1x Up direction
“Pain relief I&II”
1x Enter button
“TENS 100Hz 1”
2x Channel key #2
Favorite/ ‘Unfavorite’
3x Channel key #3
Change program level
2x Channel key #4
1x Right direction
Program description
1x Left direction
1x Channel key #1
Back to main menu
Back to program choice
Program options
Note: Final accessories may differ from the picture