10. Before Calling a Service Engineer
– 116 –
-149
Sensor Connection Error
Touch
[OK]
to close the window. Then, turn off
the power of unused sensor unit(s).
Turn on only one sensor unit when multiple
sensor units share the same body part.
Multiple sensors sharing the same BodyPart cannot be used at the same time.
Turn on the power for one sensor only that will be used.
-302
Calibration Error
Touch
[OK]
to close the window. Adjust
exposure dose (mAs) or focal distance and take
an image again.
Too Much Dose. Or collimator is used.
-303
Calibration Error
Too Little Dose. Or collimator is used.
-304
Calibration Error
• Touch
[OK]
to close the window and repeat
the exposure.
• Please contact Canon if the irradiation field
aperture is not being used.
Invalid Data. Or collimator is used.
-305
Calibration Error
Collimator may be used.
-306
QA Process Error
Touch
[OK]
to close the window, and then see
7. Processing the Image (QA Mode
Screen) (
to manually perform
QA processing.
Image Analysis Error. Adjust image process parameter manually.
-307
Calibration Error
Check that the grid has been removed, and then
retake.
When calibration is to be performed for the
CXDI-40G COMPACT, ask the service
technician to do this job.
Periodic noise is detected in acquisition data. It is necessary to remove the grid
before calibration.
-310
REX Warning
Touch
[OK]
to close the window and adjust the
dose.
REX is (detected value). It is below lower limit (setting values). Check generator
setting and exposure conditions. If exposure conditions were correct, adjust IP
parameters.
-311
REX Warning
REX is (detected value). It is over maximum limitation (setting values). Check
generator setting and exposure conditions. If exposure conditions were correct,
adjust IP parameters.
-402
Incomplete End Error
Touch
[OK]
to close the window and repeat the
exposure.
The last image in the last study was not saved completely.
-403
Incomplete End Error
• Touch
[OK]
to recover.
• Touch
[CANCEL]
to delete the image.
Do you want store captured images into internal storage in order to restore
them later?
-509
External Storage Error
Touch
[OK]
. Change the attribute of the storage
medium and try again.
Selected directory or disk for storage can only be read.
-512
External Storage Error
Touch
[OK]
to close the window and check the
setting of the external storage medium.
Cannot recognize the external storage. Set the disk properly and retry.
-513
Media Control Error
Touch
[OK]
to close the window and try again
after a while.
Operation has been canceled because the external storage or directory is busy.
Error
Code
Error Message (Upper box: Error type
Lower box: Error description)
Remedy
Содержание CXDI-1
Страница 9: ...Starting Up Shutting Down 1 Starting Up Shutting Down the CXDI ...
Страница 15: ...Exposure and Output 2 Taking an Image 3 Output Control Process Viewer ...
Страница 38: ...2 Taking an Image 30 ...
Страница 60: ...5 Viewing Multiple Images Multi View Screen 52 ...
Страница 70: ...6 Browsing Completed Studies Study List 62 ...
Страница 71: ...Processing the Image 7 Processing the Image QA Mode Screen ...
Страница 103: ...CXDI Adjustment 8 Calibration 9 Self test ...
Страница 110: ...8 Calibration 102 ...
Страница 116: ...9 Self test 108 ...
Страница 117: ...Troubleshooting 10 Before Calling a Service Engineer 11 Questions and Answers ...
Страница 136: ...11 Questions and Answers 128 ...
Страница 156: ...B Using the Rejection Reason Input Function 148 ...
Страница 162: ...C Using the Sensor Unit Attach Detach Function 154 ...
Страница 169: ......