10. Before Calling a Service Engineer
– 121 –
-533
Reject Info Output Error
Touch
[OK]
to close the window, and then see
,
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
Selected directory for reject list can only be read. Reject info cannot be saved
on the file.
-535
Reject Info Output Error
Selected directory for reject list does not exist, or attribution is incorrect. Reject
info cannot be saved on the file.
-537
Reject Info Output Error
The file format for reject list includes incorrect strings. Reject info cannot be
saved on the file.
-539
Reject Info Output Error
• Touch
[RETRY]
to store the image on the
storage again.
• To cancel save, touch
[CANCEL]
.
Error has occurred on saving reject info on the file. Do you retry it?
-540
DICOM Connect Error
• After confirming the storage is normal,
touch
[RETRY]
.
• Touch
[CANCEL]
to close the window and
then check the settings.
Transmission of the storage commitment request failed. Connection to output
equipment is not possible. Check the network or port number settings. Retry?
-541
DICOM Transfer Error
Transmission of the storage commitment request failed. An error occurred
during transmission with the output equipment. Retry?
-542
DICOM Parameter Error
Transmission of the storage commitment request failed. The DICOM output
parameter settings are invalid. Retry?
-543
DICOM Response Error
Transmission of the storage commitment request failed. The DICOM response
reported by the receiving equipment was an error. Check the receiving
equipment. Retry?
-548
Storage Commitment Request Error
Touch
[OK]
to close the window.
The storage transfer was canceled, and so the storage commitment request
was not issued.
-6000
Calibration Error
Touch
[CLOSE]
to close the window, and then
(
to perform the
calibration.
(Sensor unit ID: 1 to 4): (Elapsed time) have passed since the last calibration.
Perform calibration.
Error
Code
Error Message (Upper box: Error type
Lower box: Error description)
Remedy
Содержание CXDI-1
Страница 9: ...Starting Up Shutting Down 1 Starting Up Shutting Down the CXDI ...
Страница 15: ...Exposure and Output 2 Taking an Image 3 Output Control Process Viewer ...
Страница 38: ...2 Taking an Image 30 ...
Страница 60: ...5 Viewing Multiple Images Multi View Screen 52 ...
Страница 70: ...6 Browsing Completed Studies Study List 62 ...
Страница 71: ...Processing the Image 7 Processing the Image QA Mode Screen ...
Страница 103: ...CXDI Adjustment 8 Calibration 9 Self test ...
Страница 110: ...8 Calibration 102 ...
Страница 116: ...9 Self test 108 ...
Страница 117: ...Troubleshooting 10 Before Calling a Service Engineer 11 Questions and Answers ...
Страница 136: ...11 Questions and Answers 128 ...
Страница 156: ...B Using the Rejection Reason Input Function 148 ...
Страница 162: ...C Using the Sensor Unit Attach Detach Function 154 ...
Страница 169: ......