11. Questions and Answers
– 126 –
11.4 Exposure
Q4-1
The message QA Process Error has appeared.
A4-1
There are several causes. Typical examples and remedies are shown below:
• You have pressed the wrong exposure mode button. Select the correct exposure mode.
• Adjust the REX.
Q4-2
The density of the image cannot be adjusted as desired.
A4-2
There are several causes. Typical examples and remedies are shown below:
• Check the QA processing parameters.
(
→
• Check the IP PARAMETER for the exposure method.
(
→
Setup Guide)
• Perform calibration.
(
→
• Execute SELF TEST.
(
→
Q4-3
The image is completely white or black.
A4-3
There are several causes. Typical examples and remedies are shown below:
• The X-ray dose may be wrong, or the CXDI system may not be synchronizing with the x-ray generator.
Check the settings of the x-ray generator.
• You have pressed the wrong exposure mode button. Select the correct exposure mode.
• Adjust the REX.
Q4-4
Why is an exposure mode chosen automatically after entering a patient’s data?
A4-4
Because
AUTO-PRESS (
→
Setup Guide)
is set to
ENABLE
on the SYSTEM SETUP screen.
Содержание CXDI-1
Страница 9: ...Starting Up Shutting Down 1 Starting Up Shutting Down the CXDI ...
Страница 15: ...Exposure and Output 2 Taking an Image 3 Output Control Process Viewer ...
Страница 38: ...2 Taking an Image 30 ...
Страница 60: ...5 Viewing Multiple Images Multi View Screen 52 ...
Страница 70: ...6 Browsing Completed Studies Study List 62 ...
Страница 71: ...Processing the Image 7 Processing the Image QA Mode Screen ...
Страница 103: ...CXDI Adjustment 8 Calibration 9 Self test ...
Страница 110: ...8 Calibration 102 ...
Страница 116: ...9 Self test 108 ...
Страница 117: ...Troubleshooting 10 Before Calling a Service Engineer 11 Questions and Answers ...
Страница 136: ...11 Questions and Answers 128 ...
Страница 156: ...B Using the Rejection Reason Input Function 148 ...
Страница 162: ...C Using the Sensor Unit Attach Detach Function 154 ...
Страница 169: ......