11. Questions and Answers
– 125 –
11.3 Settings
Q3-1
A new exposure mode button cannot be made.
A3-1
To create a new exposure method, copy the exposure method used before and then edit it.
(
→
Setup
Guide)
This eliminates the need to enter the parameters from the beginning.
Q3-2
What is a bucky msec?
A3-2
It is a measure of the time for linking the movable grid with the X-ray irradiation. When it is not set
properly, vertical lines may appear on the image.
Q3-3
What is the difference between the fixed ROI and phototimer?
A3-3
In the
Fixed ROI
settings, the adjustment range for
LUT processing (
→
is set. By setting the
phototimer
, the position of the phototimer installed in the CXDI is set.
Q3-4
What is an X-ray monitor?
A3-4
It is used to detect the completion of X-ray irradiation and judge the timing for image generation.
Normally, this cannot be changed.
(
→
Setup Guide)
Содержание CXDI-1
Страница 9: ...Starting Up Shutting Down 1 Starting Up Shutting Down the CXDI ...
Страница 15: ...Exposure and Output 2 Taking an Image 3 Output Control Process Viewer ...
Страница 38: ...2 Taking an Image 30 ...
Страница 60: ...5 Viewing Multiple Images Multi View Screen 52 ...
Страница 70: ...6 Browsing Completed Studies Study List 62 ...
Страница 71: ...Processing the Image 7 Processing the Image QA Mode Screen ...
Страница 103: ...CXDI Adjustment 8 Calibration 9 Self test ...
Страница 110: ...8 Calibration 102 ...
Страница 116: ...9 Self test 108 ...
Страница 117: ...Troubleshooting 10 Before Calling a Service Engineer 11 Questions and Answers ...
Страница 136: ...11 Questions and Answers 128 ...
Страница 156: ...B Using the Rejection Reason Input Function 148 ...
Страница 162: ...C Using the Sensor Unit Attach Detach Function 154 ...
Страница 169: ......