Continued on the following page >>
10. Before Calling a Service Engineer
– 119 –
10.4 Warning
Error
Code
Error Message (Upper box: Error type
Lower box: Error description)
Remedy
-9
SMART Error
Touch
[OK]
to close the window, and then see
,
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
An error is detected in HDD.
It's possible that the HDD(1 or 2) will break down. Call a service person.
-10
HDD Communication Error
Cannot communicate with HDD.
Mirroring function will be disabled since the HDD(1 or 2) is not available. Call a
service person.
-11
SMART Error
An error is detected in HDD.
It's possible that the HDD(1 or 2) will break down and the data will be lost. Call
a service person.
-150
Sensor Unit: SelfCheck Error
Detected an Error.
-151
Sensor Unit: Flash ROM Error
Detected Flash ROM Error. Call a service engineer.
-152
Sensor Unit: Sequence Error
Touch
[OK]
to close the window. As image
quality cannot be guaranteed, take an image
again.
This image may have a problem.
-153
Sensor Unit: Temperature Warning
Touch
[OK]
to close the dialogue window.
Lower the temperature of the room where the
sensor unit is installed, or wait until the
temperature in the sensor unit drops.
The internal temperature of the Sensor Unit is too high!
Do not leave the Sensor Unit in READY condition.
-160
Capture Board Error
Touch
[OK]
to close the window, and then see
,
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
Detected Capture Board Error.
-161
Capture Board Error
Capture Board Error.
-162
Capture Board: Sequence Error
X-ray monitor in sensor unit failed to detect the beginning of irradiation.
-163
Capture Board: Sequence Error
X-ray monitor in sensor unit failed to detect the end of irradiation.
-164
Image quality attention
Touch
[OK]
to close the window. As image
quality cannot be guaranteed, take an image
again.
CXDI has stopped irradiation. Check image.
-502
DICOM Connect Error
• After confirming that the network is normal,
touch
[RETRY]
.
• Touch
[CANCEL]
to close the window, and
then see
Setup Guide
and set the
parameters.
• If running the software offline, check that
the output status of the process viewer is
[IDLE]
. If output status is
[SEND]
, change it
to
[IDLE]
.
Cannot connect to target. Check network or port number setting.
(The following message is added when transferring all data at once.)
If operating offline, press the Close button and check the output status of the
Process Viewer.
-503
DICOM Transfer Error
Error has occurred during association.
Содержание CXDI-1
Страница 9: ...Starting Up Shutting Down 1 Starting Up Shutting Down the CXDI ...
Страница 15: ...Exposure and Output 2 Taking an Image 3 Output Control Process Viewer ...
Страница 38: ...2 Taking an Image 30 ...
Страница 60: ...5 Viewing Multiple Images Multi View Screen 52 ...
Страница 70: ...6 Browsing Completed Studies Study List 62 ...
Страница 71: ...Processing the Image 7 Processing the Image QA Mode Screen ...
Страница 103: ...CXDI Adjustment 8 Calibration 9 Self test ...
Страница 110: ...8 Calibration 102 ...
Страница 116: ...9 Self test 108 ...
Страница 117: ...Troubleshooting 10 Before Calling a Service Engineer 11 Questions and Answers ...
Страница 136: ...11 Questions and Answers 128 ...
Страница 156: ...B Using the Rejection Reason Input Function 148 ...
Страница 162: ...C Using the Sensor Unit Attach Detach Function 154 ...
Страница 169: ......