BÜCHI Labortechnik AG
Installation | 5
Operation Manual ProxiMate™
25/94
Navigation path
➔
➔
[Calibrate BCV]
Precondition:
R
The instrument is in administrator mode. See Chapter 6.5 "Log in administrator
mode", page 39
R
A sample presentation is defined.
R
The instrument is stabilized with at least 2 hours continuous running.
u
Navigate to the action
[Calibrate BCV]
via the navigation path.
ð
The instrument shows a dialog box with mandatory settings.
5
4
3
2
1
Fig. 10: Dialog box
Calibrate BCV
1
Measurement view
2
External Reference
3
Progress (view)
4
Measurement mode
5
Sample presentation ID
u
Select the name of the sample presentation you wish to calibrate.
u
Enter the corresponding values.
u
Place the white reference or empty sample presentation method in the correct po-
sition.
u
Tap the button
[Start]
.
u
Follow instructions on the display during the calibration.
u
Confirm the instructions by tapping the
[OK]
button.
5.6.4
Import external reference data
NOTE
The location of the destination folder is fixed. See Chapter 10.2 "Folder locations",
page 89
Navigation path
➔
➔
[Import External Refer-
ence Data]
Precondition:
R
The reference data file is saved on the instrument.
u
Navigate to the action
[Import External Reference Data]
via the navigation path.
ð
The display shows the dialog box
Import External Reference Data
.