7 | Cleaning and servicing
BÜCHI Labortechnik AG
72/94
Operation Manual ProxiMate™
Component
Action
Frequency
Display
u
Wipe down the display with a damp
cloth. See Chapter 7.7 "Cleaning the
display", page 85
Weekly
Sample carrier slid-
ers
u
Check that the sample carrier sliders
do not show excessive wear.
Monthly
Control panel
u
Carry out an Advanced System Test.
See Chapter 7.5.3 "Carrying out an
Advanced System Test", page 84
Monthly
Desiccant cartridge
u
Check function of the desiccant car-
tridge.Chapter 7.3 "Checking the des-
iccant cartridge", page 80
u
If necessary change the desiccant
cartridge. See Chapter 7.4 "Changing
the desiccant cartridge", page 82
Yearly
Control panel
u
Carry out a Comprehensive System
Test. See Chapter 7.5.2 "Carrying out
a Comprehensive System Test",
page 84
Yearly
7.2
Changing the lamps
NOTICE
Risk of broken lamp
Touching the bulb or the reflector with hands can cause lamp damage.
u
Do not touch the bulb with fingers.
u
Touch the lamp on the indicated areas.
7.2.1
Changing the down view lamp
u
Disconnect the power supply to the
instrument.
u
Wait 15 min. for lamp to cool.