6 | Operation
BÜCHI Labortechnik AG
60/94
Operation Manual ProxiMate™
u
Tap the button
[OK]
.
ð
The dialog box closes.
ð
The value for the Decimal places is saved.
6.8.16
Changing the Unit of a property
NOTE
Special characters
4
Tab the button
[Shift]
on the alphanumeric input box.
ð
The alphanumeric input box shows the available special characters.
Navigation path
➔
Precondition:
R
The instrument is in administrator mode. See Chapter 6.5 "Log in administrator
mode", page 39
u
Navigate to the
[Application]
menu via the navigation path.
u
Tap the function
[Edit]
on the function bar.
ð
The display highlights the function
[Edit]
in green.
u
Tap the name of the application that you wish to edit.
ð
The display highlights the application in green.
u
Tap the name of the property that you wish to edit.
ð
The display shows the view
Property
.
u
Tap the action
[Unit]
.
ð
The display shows a dialog with an alphanumeric input box.
u
Enter a unit for the property.
u
Tap the button
[OK]
.
ð
The dialog box closes.
ð
The Unit is saved.
6.8.17
Changing the Bias of a property
A bias is a constant value.
This value is added to the result of a prediction to correct a constant deviation be-
tween predicted values and reference values.
Proceedings:
Navigation path
➔
Precondition:
R
The instrument is in administrator mode. See Chapter 6.5 "Log in administrator
mode", page 39
u
Navigate to the
[Application]
menu via the navigation path.
u
Tap the function
[Edit]
on the function bar.
ð
The display highlights the function
[Edit]
in green.