BÜCHI Labortechnik AG
Cleaning and servicing | 7
Operation Manual ProxiMate™
85/94
Navigation path
➔
➔
[Backup NIRWise Data]
Precondition:
R
The instrument is in administrator mode. See Chapter 6.5 "Log in administrator
mode", page 39
u
Navigate to the action
[Backup NIRWise Data]
via the navigation path.
ð
The display shows a dialog with values that can be saved.
u
Select the values according to your needs.
u
Tap the button
[Start]
.
ð
The instrument creates a .zip file with the selected data.
u
Save the data to an external data store.
7.7
Cleaning the display
The cleaning mode switches off the touch function of the touchscreen.
Navigation path
➔
➔
[Cleaning Mode]
u
Navigate to the action
[Cleaning Mode]
via the navigation path.
ð
The touch function of the display is locked.
ð
The display shows a dialog box with a numeric input box.
u
Wipe down the display with a damp cloth.
u
Enter the displayed secure code.
u
Tap the button
[OK]
.
ð
The dialog box closes.
ð
The touch function of the display is unlocked.