7 | Cleaning and servicing
BÜCHI Labortechnik AG
76/94
Operation Manual ProxiMate™
u
Attach the cover lid to the instrument.
u
Change the desiccant cartridge. See
Chapter 7.4 "Changing the desiccant
cartridge", page 82
u
Confirm lamp replacement. See
Chapter 7.2.3 "Confirm Lamp Re-
placement", page 80
u
Carry out a baseline correction vector
calibration. See Chapter 5.6.3 "Cali-
brating a Baseline Correction Vector
(BCV)", page 24
7.2.2
Changing the up view lamp
u
Disconnect the power supply to the
instrument.
u
Wait 15 min. for lamp to cool.
u
Turn the instrument around.
u
Make sure that the instrument cannot
tip over during the lamp replacement
procedure.