Proceedings:
Navigation path
➔
Precondition:
R
The instrument is in administrator mode. See Chapter 6.5 "Log in administrator
mode", page 39
u
Navigate to the
[Application]
menu via the navigation path.
u
Tap the function
[Edit]
on the function bar.
ð
The display highlights the function
[Edit]
in green.
u
Tap the name of the application that you wish to edit.
ð
The display highlights the application in green.
u
Tap the action
[Rotation]
.
ð
The display shows a dialog box with the selectable values.
u
Select a value.
u
Tap the button
[OK]
.
ð
The dialog box closes.
ð
The setting is saved.
6.6.9
Changing the Measurement duration of an application
The default measurement time (15 seconds) allows the sample to complete one rota-
tion. Shorter measurement time does not allow a complete rotation of the sample.
Proceedings:
Navigation path
➔
Precondition:
R
The instrument is in administrator mode. See Chapter 6.5 "Log in administrator
mode", page 39
u
Navigate to the
[Application]
menu via the navigation path.
u
Tap the function
[Edit]
on the function bar.
ð
The display highlights the function
[Edit]
in green.
u
Tap the name of the application that you wish to edit.
ð
The display highlights the application in green.
u
Tap the action
[Measurement duration]
.
ð
The display shows a dialog box with a numeric input box.
u
Enter the value in the numeric input box.
u
Tap the button
[OK]
.
ð
The dialog box closes.
ð
The value for the duration is saved.
6.6.10
Enter a Standard Operating Procedure (SOP) for an application
Each Application has an optional Standard Operating Procedure (SOP) associated
with it. A Standard Operating Procedure SOP shows instructions for the user to follow
during measurement.