BÜCHI Labortechnik AG
Cleaning and servicing | 7
Operation Manual ProxiMate™
71/94
7
Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this sec-
tion.
Any servicing and repair work which involves opening up the casing may only be car-
ried out by BUCHI service technicians.
u
Use only genuine BUCHI consumables and spare parts in order to ensure correct
operation of the device and preserve the warranty.
7.1
Regular maintenance work
NOTE
If special cleaning processes are necessary contact BUCHI Customer Service.
Component
Action
Frequency
Sample vessel
u
Wipe down the sample containers
with a damp cloth.
u
If heavily soiled:
use mild detergent
rinse with clean water
wipe dry with a lint free cloth
Daily
Glass up view
u
Wipe down the glass window with a
damp cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Daily
Glass down view
u
Wipe down the glass window with a
damp cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Daily
Sample presenta-
tion
u
Wipe down the sample presentation
area with a damp cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Daily
Data
u
Perform a data backup. See Chap-
ter 7.6 "Carrying out a data backup",
page 84
Weekly
Casing
u
Wipe down the casing with a damp
cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Weekly
Cooling fins
u
Remove dust and foreign objects
from the cooling fins using com-
pressed air or a vacuum cleaner.
Weekly
Control panel
u
Carry out an Baseline Correction Vec-
tor test. See Chapter 7.5.1 "Carrying
out a Baseline Correction Vector
test", page 83
Weekly