7 | Cleaning and servicing
BÜCHI Labortechnik AG
84/94
Operation Manual ProxiMate™
u
Navigate to the action
[Test BCV]
via the navigation path.
ð
The display shows the dialog
Test BCV
.
u
Select the name of the sample presentation you wish to test.
u
Tap the button
[Start]
.
u
Follow instructions on the display during the test.
u
Confirm the instructions by tapping the
[OK]
button.
u
Once the test is complete, the instrument records the results in a report.
ð
The display shows Baseline Correction Tests is completed.
7.5.2
Carrying out a Comprehensive System Test
Navigation path
➔
➔
[Extended System Test]
Precondition:
R
The instrument is in administrator mode. See Chapter 6.5 "Log in administrator
mode", page 39
R
External reference data for all standards are loaded.
R
Performance Test Standards Kit is available.
u
Navigate to the action
[Extended System Test]
via the navigation path.
ð
The display shows the dialog
Extended System Test
.
u
Select check box
[Comprehensive System Test]
u
Tap the button
[Start]
.
u
Follow instructions on the display during the test.
u
Confirm the instructions by tapping the
[OK]
button.
u
Once the test is complete, the instrument records the results in a report.
ð
The display shows Comprehensive System Tests is completed.
7.5.3
Carrying out an Advanced System Test
Navigation path
➔
➔
[Extended System Test]
Precondition:
R
The instrument is in administrator mode. See Chapter 6.5 "Log in administrator
mode", page 39
u
Navigate to the action
[Extended System Test]
via the navigation path.
ð
The display shows the dialog
Extended System Test
.
u
Select check box
[Advanced System Test]
u
Tap the button
[Start]
.
ð
The instrument starts the test.
u
Once the test is complete, the instrument records the results in a report.
ð
The display shows Advanced System Tests is completed.
7.6
Carrying out a data backup
NOTE
The location of the destination folder is fixed. See Chapter 10.2 "Folder locations",
page 89