Reference Time
: The memory cycle is synchronized using this value.
Data Log Interval in Alarm Condition
: In alarm condition substantially more data
needs to be stored in the log file. Default is 3 minutes. The system switches auto-
matically to a shortened log interval when an emergency occurs.
3.
Clear Event / Data Log
In order to delete the entire data or event log file, you must use the Clear Data Log
button.
i
Warning: Using the Clear Data Log button will delete all the data in the file!
It is advisable that you backup the log files on an external storage device (e.g. CD-ROM)
before using the Clear Data Log function. To do this you can use the Export Data and
Event Log button and/or save the complete log.mdb and log.ldb file from the database
directory.
12.3.3.3
He-Refill / He Measurement
This settings window is used to adjust the properties of the helium refill function (refer to
the section
Figure 12.19Tools - Options - He-Refill / He Measurement
The helium refill time setting allows you to set the amount of time that the selected
alarms will be deactivated during the helium refill. These alarm parameters are set in the
Sensor Channels section under the Monitoring System settings (see also
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......