39
Z31823_00_05
6
Alarm System
The alarm system is an important part of the BMPC/2 and is needed for the supervision
of the magnet system. The alarm system uses an analog telecommunication network,
based on an analog fax modem (with its own modem controller), and normally is avail-
able for use worldwide.
6.1
Alarm System Function
When the BMPC/2 is working in the Monitor & Alarm mode, all system parameters are
continuously monitored and compared with their limit values. When a parameter
exceeds or falls below a given alarm limit, an alarm cycle will be started.
In the following example the alarm cycle was triggered because of a low helium flow.
The status field shows that an alarm call was activated and the resulting events logged
in the Event Log section.
Figure 6.1 Present Values/Events in Alarm Mode
1. Helium Flow Alarm is enabled
2. Alarm Fax is sent to Bruker BioSpin
3. Confirmed by „Mustermann, Klaus“
4. System called Mustermann, Klaus
5. Helium Flow is below alarm limit
6. System must be in Monitor & Alarm Mode
7. System status is now Alarm Call
7
6
5
5
4
3
2
1
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......