Setup and Diagnostics Tool
10.4
Calibrate Temperature Linearization Function
This function is used to recalibrate the temperature sensors for T1 or T2. To use this
function you need both calibration connectors BMPC/2 T1/T2 Calibration Connector
High (P/N HZ13703) and BMPC/2 T1/T2 Calibration Connector Low (P/N HZ13704).
Start the calibration by selecting the Calibrate Temperature Linearization Function.
Figure 10.4 Setup Tool - Calibrate Temperature
1. First input the values of both calibration connectors in the fields shown in the figure
above and press the
Save Calibration Values on Disk
button. These values are
printed on a small label on the calibration connector. Once entered the values will
be stored on the hard drive for future use.
2. Press the button
Calibrate T1 & T2
and follow the instructions that will be displayed.
i
Store the calibration connectors very carefully. Without these parts it is not possible to
recalibrate the BMPC/2.
Values from HZ13703
Values from HZ13704
Start Calibration Cycle
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......