Alarm System
6.4
Alarm Groups
6.4.1
Internal Group
Only company internal phone numbers should be assigned to the „Internal Group”. Typi-
cally, during the week, the likelihood of reaching someone in the „Internal Group” is high
so a response to the alarm should be very fast. Depending of the configuration it may
need more than one loop through the „Internal Group” before anyone acknowledges the
alarm call using the „Confirmation code”. (see also
).
6.4.2
External Group
The „External Group“ list normally stores the private telephone numbers or cell phone
number for the individuals responsible for the system. If no one from the internal group
acknowledges the alarm, the system will start calling all the people in the „External
Group“.
6.4.3
Fax Group
When the fax group is enabled, the system will send an alarm fax to all phone numbers
in the „Fax Group” during the alarm sequence.
When the person who has assumed responsibility for the system (by acknowledging the
alarm call) switches the system on site to „OFF“ mode, a confirming fax will be sent to all
alarm fax recipients to show that the magnet system is under control.
6.4.4
E-mail Group
This group is optional. When this option is enabled the system will send an e-mail to the
E-mail Group.
For information on setting up e-mail for notification, see
6.5
Stopping the Alarm Sequence
To stop an alarm sequence on site, switch the system to „OFF” mode. Once the problem
has been fixed, switch the system to „Monitor & Alarm” mode.
i
The behavior of the system in the case of an alarm is dependent on the settings.
The behavior described above is the case of the default settings (delivery state).
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......