70
Z31823_00_05
Troubleshooting
11.1.1
Description of Error Messages
The following section contains a description of some of the common error messages that
might be displayed, as well as a solution to the problem.
11.1.1.1
Limit Data P1 failed
It was not possible to read the limit data for the pressure P1 from the EEProm, thus
supervision of the Pressure P1 is not possible.
Solution:
Using the function „Read from Backup Database“ from the menu
Tools -
) you are able to read the limit val-
ues stored on the PC’s hard disk.
To solve this problem you will need to reload the values from the backup database to the
BMPC/2 Control Unit using the „Write Values“ function.
11.1.1.2
Limit Data T1 Failed
It was not possible to read the limit data for the temperature T1 from the EEProm, thus
supervision of the temperature T1 is not possible.
Solution
: Reload the values from the backup database to the BMPC/2 as shown for
Limit Data P1 fail.
11.1.1.3
Limit Data T2 Failed
It was not possible to read the limit data for the temperature T2 from the EEProm, thus
supervision of the temperature T2 is not possible.
Solution
: Reload the values from the backup database to the BMPC/2 as shown for
Limit Data P1 fail.
11.1.1.4
Limit Data T5 Failed
It was not possible to read the limit data for the temperature T5 from the EEProm, thus
supervision of the temperature T5 is not possible.
Solution
: Reload the values from the backup database to the BMPC/2 as shown for
Limit Data P1 fail.
11.1.1.5
Limit Data Flow Failed
It was not possible to read the limit data for the helium flow Fl from the EEProm, thus
supervision of the helium flow Fl is not possible.
Solution
: Reload the values from the backup database to the BMPC/2 as shown for
Limit Data P1 fail.
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......