100
Z31823_00_05
Configuration (administrator)
12.3.2
Pump Control
12.3.2.1
Limit Values
The Pump Control option setup window allows you to set the various limit values for site
conditions. Default values are normally already assigned during the magnet setup wiz-
ard, thus an adjustment is only needed in rare cases.
When you want to change a limit value, first select the parameter, then enter the new
upper and lower limit values. To transfer the new values to the BMPC/2 press the Write
Values button.
Figure 12.14Tools - Options - Limit Values
•
Read Values: Pressing this button will result in the display of the actual values from
the BMPC/2. This is useful when you have changed the values, but not written them
to the unit and want to reload the actual values.
NOTICE
Changing the limit values
Changing the limit values may result in the system reacting differently then expected
and may delay the triggering of alarms.
Changes should only be made by experienced personnel that understand the
implications of these changes.
Selected
Parameter
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......