61
Z31823_00_05
Hardware Overview
8.6.4.2
Connector X6 UPS Status Interface
Figure 8.14 Line Control Unit Connector X6
i
The cable HZ13979 must be connected to connector X6 or it is not possible to switch on
the console output and to detect commercial power failures.
Connection to UPS
Status Interface
Pin Number
Pin Function
1
Not connected
2
Not connected
3
General alarm (normally opened)
4
On commercial AC (normally opened)
5
Ground
6
UPS on bypass (normally opened)
7
Battery low (normally opened)
8
UPS present (normally closed)
9
Not connected
Table 8.6 Pin Count Connector X6