110
Z31823_00_05
Configuration (administrator)
In the setup category LIMIT VALUES (see chapter "Limit Values") you will find among
others the following two entries used to setup the limit values for use with the HLMU:
- Helium Level (HLMU)
- Nitrogen Level (HLMU)
It is recommended to use the given default parameters (Normal = 70, Warning = 60,
Alarm = 50).
Note: If you are using the HLMU for helium (nitrogen) level measurements, the channel
"Helium Level" that is normally used, must be disabled (see
), and the two channels Helium Level (HLMU) and Nitrogen Level (HLMU) must be
enabled.
When you now start the Monitoring Mode/Monitoring & Alarm Mode the system will first
try to establish a connection to the HLMU. If it fails, the system will not start and an entry
will be logged in the Event Log. When the connection is successful, the system will begin
continuously reading the level(s) values from the HLMU.
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......