79
Z31823_00_05
Troubleshooting
11.3.2
Version Check Error
The BMPC/2 control program requires a firmware version of VX.Y or higher. When this is
not the case, the communication between the PC and the BMPC/2 control unit will not
function properly. Refer to the section
"Downloading New Firmware" on page 66
information on how to download the latest version of firmware.
11.3.2.1
COM Port Error
When a COM Port error occurs, it means that the communications port used for commu-
nication with the BMPC/2 cannot be used.
Either the port is in use from another program (e.g. SBS terminal or the BMPC/2 Setup
Tool), or the cable connection to the control unit is not functioning correctly.
The standard configuration for the COM Port is COM1. Please check this configuration
in
Tools - Options - General - Communication Setting
. Also refer to section
.
11.4
Other Errors
11.4.1
When the Value of a Parameter is Displayed as NAN
When the value of a parameter is displayed as NAN (Not A Number), it means that the
value of a sensor that is connected to the CAN bus cannot be read. This can lead to var-
ious consequences.
Solution:
1. Try to re-initialize the CAN bus by doing a software reset of the BMPC control unit
(see
).
2. If this is not successful, try to turn the control unit on and off by opening and closing
the Q10 safety fuse on the BMPC/2. This is accessible by opening the rear side,
upper cover of the BMPC. The Q10 safety fuse is on the upper left-hand side.
When this still not solves the problem, check the CAN bus wiring (see
).
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......