Software
5.4.4
He Fill Mode
The He Fill mode is necessary because the T1 and/or T2 parameters may briefly fall
below the alarm threshold during the helium refill process. In order to avoid an alarm
from being triggered the He Fill mode can be used to suppress the T1/T2 alarm for up to
six hours.
The P2 pressure control heater is also disabled for one hour, as this is not efficient when
the magnet is open and thus would only lead to unnecessary helium consumption.
When the He Fill mode is activated it is indicated in the status bar together with the
length of fill time remaining.
Figure 5.16 He Fill Mode 292 Minutes
The suppressed alarms are represented in the Present Values/Event display as indi-
cated in the figure below.
Figure 5.17 Alarm Temporarily Disabled
Once the helium refill is finished and the remaining time was expired, the program will
automatically leave He Fill mode and return to monitoring mode.
For information on configuration the He Fill mode read
"He-Refill / He Measurement" on
Alarm temporarily
disabled
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......