Power Analysis
193
19.8
Turn on/Turn Off
Turn on analysis determines the time taken for the power supply to reach 90% of its steady-state output. Turn off analysis
determines the time taken for the power supply to fall to 10% of its maximum output voltage.
19.8.1
Turn On/Turn Off Input Setup
1.
Touch
Input Setup
to display the signal menu shown in figure
2.
Touch
Output Voltage
to set the input voltage source.
3.
Touch
Input Type
to toggle the input type between
AC
and
DC
.
4.
Touch
Duration
and turn the
Universal Knob
or use the virtual numeric
keypad to set the duration time.
–
the range is 5 ns 20 s. The oscilloscope will set the timebase according
to the value (duration / 10)
5.
Touch
Max V in (RMS)
and turn the
Universal Knob
or use the virtual
numeric keypad to set the maximum effective input voltage (Vrms).
–
The range is 1 V to 1 kV.
–
The oscilloscope will set the vertical scale of the voltage channel to maxi-
mum input voltage / 7 and the trigger level to maximum input voltage /
10.
6.
Touch
Stable V out
and turn the
Universal Knob
or use the virtual
numeric keypad to set the stable output voltage value.
–
The range is -30 V to 30 V.
–
The oscilloscope will set the vertical scale of the output voltage channel
to stable output voltage / 6
7.
Touch
Connection Guide
to display a diagram illustrating the procedure
setup. See figure
Figure 19.19
Input
Setup Menu