Serial Trigger and Decode
109
WS
•
Use the touchscreen to select
WS
and select the source the signal is connected to.
•
Use the touchscreen to select
Threshold
and input the desired threshold
–
The threshold voltage level is for decoding, and it will be regarded as the trig-
ger voltage level when the trigger type is set to serial.
–
The threshold level line only appears when adjusting the threshold level.
•
Use the touch screen to select either
Low
or
High
Left CH.
–
Low
: Select left channel when WS is low and right channel when WS is high.
–
High
: Select the right channel when WS is low and left channel when WS is
high.
•
Use the touchscreen to select
Data
the
Source Setting
window will appear.
–
Select the channel that will provide the data signal.
–
Use the touchscreen to select
Threshold
and input the desired threshold.
–
Touch
option to return previous menu.
Protocol Config
To configure the protocol touch
1.
The
I2S CONFIG
menu will be displayed. (Figure
2.
Use the touchscreen to select
Audio Variant
. The following variants are available:
–
Audio-I2S
–
Audio-LJ
–
Audio-RJ
3.
Use the touchscreen to select
Start Bit
and input the starting bit.
4.
Use the touchscreen to select
Data Bits
and enter the word size. (4 to 32 bits)
–
The word size of the device under test must be matched.
5.
Use the touchscreen to select
LSB
or
MSB
to set the bit order.
6.
Use the touchscreen to select
Annotate
. The following option are available:
–
All
–
Left
–
Right
7.
Touch
option to return previous menu.
Figure 10.41
I2S CON-
FIG
Содержание 2560B Series
Страница 1: ......
Страница 51: ...Digital Channels 51 Figure 8 9 Digital Bus...
Страница 127: ...Serial Trigger and Decode 127 Figure 10 63 Manchester Bus...
Страница 168: ...Reference Waveform 168 Figure 14 1 Recall REFB...