User's Manual
•
5-
3
the guides rails located in the enclosure and the entire transilluminator module will
become free from the enclosure.
Release lever for slides
Transilluminator
Transilluminator without the cover
Figure 5.1
6. Using the Phillips screwdriver, remove the cover assembly from the transilluminator by
removing four screws. There are two screws on each side of the cover.
7. Using the 5/16" socket wrench, remove the four hex standoffs that hold the baffles on the
lamps of the transilluminator and then gently lift off the baffles exposing the
transilluminator lamps.
Note: Be careful and do not bend the baffles because bending the baffles could lead to deterioration of
uniformity of your UV transilluminator.
8. Remove lamps by holding the metal part of the lamps with the tip of your fingers and
turning them in a rotating movement and by pulling them out of the socket. (See Fig 5.2)
Remove hex standoffs (X4)
Remove the baffle exposing the lamp
Hold at the metal ends of the
lamps and turn to remove
Figure 5.2
9. Install the replacement lamp by inserting its ends in the sockets and rotating until it clicks
into place.
10. Now reinstall the baffles on the lamps by making sure that they are seated correctly.
11. Reinstall the cover on the transilluminator module. Replace the screws.
12. Carefully reinstall the transilluminator assembly back onto the slides by carefully lining up
the slides into the guides and then gently push in all the way.
NOTE: There is no need to push down or up on the levers, they are only used to remove the
transilluminator assembly.