User's Manual
•
3-13
Figure 3.14b Figure 3.14c
6. Next attach the epi white module to the epi UV module by locking in the tongue and
groove located at the outer top edges of each module. (See figure 3.14c)
7. Now secure the epi white module to the epi UV module using the latch and screw
mechanism located at the front plate of the modules. See Figure 3.15
Figure 3.15
8. Next connect the cable from the epi white module to the appropriate connector on
the ballast assembly.
9. Re-install the access panel to the ballast panel.
NOTE: The plugs for each illumination module are keyed and cannot be interchanged.
Each plug will snap together only to the appropriate socket.
3.2.7 Electrical and Communication Connections
Power
The power entry module of the VersaDoc system is configured for operation at 100/120
VAC, 60 Hz or 220-260 VAC, 50 Hz. After checking that the system’s power switch is
turned off, insert an approved power cord into the power entry module located on the
bottom right of rear panel of the unit. See Figure 3.16