BeneVision TM80 Telemetry Monitor Operator’s Manual
7 - 27
■
When the 5-lead ECG leadwires are used, seven ST values (ST-I, ST-II, ST-III, ST-aVR, ST-
aVL, ST-aVF, ST-V) are displayed in the ST digital area.
■
When the 6-lead ECG leadwires are used, eight ST values (ST-I, ST-II, ST-III, ST-aVR, ST-
aVL, ST-aVF, ST-V
a, ST-Vb
) are displayed in the ST digital area.
This example shows the ST digital area when the 6-lead ECG leadwires are used. Your
screen may look slightly different:
1.
Parameter label
2.
ST alarm off symbol
3.
Lead label
4.
ST unit
5.
ST numerics: a positive value indicates ST segment elevation, and a negative value
indicates ST segment depression.
7.8.4 Configuring ST Alarm Settings
To configure ST alarm settings, follow this procedure:
1.
Tap the ST digital area to enter the
ST
menu.
2.
Tap
ST Alarm Setup
.
3.
In the
Setup
section, tap
ST Alarm Mode
and then select the desired mode:
◆
Absolute
:
you can separately set the alarm properties for each ST alarm. After
selecting this option, the Auto Limits option is displayed where you can initiate
auto adjustment of alarm limits.
◆
Relative
: it is the default option. You can set the alarm properties for
ST Single
and
ST Dual
alarms.
4.
In the
Alarms
section, select the options described in the following table.
1
3
4
)
2
5
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...