12 - 4
BeneVision TM80 Telemetry Monitor Operator’s Manual
3.
Tap the
icon to exit the
Alarms
menu.
WARNING
•
When Minimum Alarm Volume is set to Off and Sounds is set to 0, alarm sound
is turned off, the telemetry monitor will not enunciate alarms when they occur.
Be careful when turning off the alarm sound.
•
Do not rely exclusively on the audible alarm system for monitoring. Adjustment
of alarm volume to a low level may result in a hazard to the patient.
Reminder
Interval
Allows an authorized user to configure
the intervals between the alarm tones.
1 min, 2 min, 3 min,
5 min
,
10 min
Technical Alarm Priority
ECG Lead Off
Allows an authorized user to configure
the alarm priority.
Low
, Medium, High
SpO
2
Sensor Off
Allows an authorized user to configure
the alarm priority.
Low
, Medium, High
Section &
Options
Description
Settings*
*
The factory default settings are in bold.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...