9 - 10
BeneVision TM80 Telemetry Monitor Operator’s Manual
2.
Tap the
icon to exit the
SpO
2
menu.
9.6
SpO
2
Measurement
9.6.1 Identifying SpO
2
Modules
To identify which SpO
2
module you are using, see the company logo on the SpO
2
module.
■
Masimo SpO
2
module is with a logo of Masimo SET.
■
Nellcor SpO
2
module is with a logo of Nellcor.
Options
Description
Settings*
SpO
2
Alarm Setup
Configures whether to switch on SpO
2
and
SpO
2
Desat alarms.
SpO
2
and
Desat
default
to be switched on.
After tapping
SpO
2
or
Desat
you can
configure alarm limits and alarm priority.
For SpO
2
:
Alarm switch:
On
, Off
Alarm limit range:
0% to 100%;
The default alarm upper
limit is
100
%. The default
alarm low limit is 90%.
Alarm Priority:
Med
, High
For Desat:
Alarm switch:
On
, Off
Alarm limit range:
Lower alarm limit is not
higher than the SpO
2
lower
alarm limit. The default
alarm low limit is
80%.
Alarm Priority:
High
NIBP Simul
When monitoring SpO
2
and NIBP on the same
limb simultaneously, you can switch on NIBP
Simul to lock the
SpO
2
alarm status until the
NIBP measurement ends. If you switch off
NIBP Simul
, low perfusion caused by NIBP
measurement may lead to inaccurate SpO
2
readings and therefore cause false
physiological alarms.
On,
Off
*
The factory default settings are in bold.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...