BeneVision TM80 Telemetry Monitor Operator’s Manual
10 - 9
SYS
Extreme
Configures whether to trigger
the extreme NIBP systolic
pressure alarm.
On,
Off
Alarm high limit:
• For adult:
Range: (SYS high limit + 5) to 290
Default value:
175
• For pediatric:
Range: (SYS high limit +5) to 240
Default:
130
Alarm low limit:
• For adult:
Range: 25 to (SYS low limit - 5)
Default value:
75
• For pediatric:
Range: 25 to (SYS low limit - 5)
Default:
60
Alarm priority:
High
DIA
Extreme
Configures whether to trigger
the extreme NIBP diastolic
pressure alarm.
On,
Off
Alarm high limit:
• For adult:
Range: (DIA high limit + 5) to 250
Default value:
105
• For pediatric:
Range: (DIA high limit + 5) to 200
Default:
80
Alarm low limit:
• For adult:
Range: 10 to (DIA low limit - 5)
Default value:
35
• For pediatric:
Range: 10 to (DIA low limit - 5)
Default:
30
Alarm priority: High
Options
Description
Settings*
*
The factory default settings are in bold.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...