13 - 10
BeneVision TM80 Telemetry Monitor Operator’s Manual
If both V-Tach PVCs and V-Tach Rate are greater than or equal to the limits, a V-Tach alarm is
generated.
■
If consecutive PVCs is lower than the V-Tach PVCs limit (6) but greater than 2, and the
Vent rate is greater or equal to the V-Tach Rate limit (130), a Nonsus V-Tach alarm is
generated.
■
If consecutive PVCs is greater than or equal to the V-Brady PVCs limit (5), and the Vent
rate is lower than the V-Brady limit (40), a Vent Brady alarm is generated.
■
If both the V-Brady PVCs and V-Brady Rate are lower than the limits, but V-Brady PVCs
is greater than 2, a Run PVCs alarm is generated.
■
If the V-Brady PVCs and V-Brady Rate are greater than or equal to limits, but the Vent
rate is is lower than V-Tach Rate (130), a Vent Rhythm alarm is generated.
13.7.2 Configuring the Arrhythmia Shield Time
To configure arrhythmia shield time, follow this procedure:
1.
Select the system menu area
in the upper left corner of the CMS screen.
2.
S
elect
System Setup
to access the
System Setup
menu.
3.
Select the
Telemetry
tab.
4.
Select the
Alarm
Tab.
5.
Set
Arrhy Shield Time
. When it is set to
0
, it indicates that the alarm alarm shield
function is disabled.
13.8 Locating the Telemetry Monitor
If the AP information of the TM80 has been imported into the CentralStation, you can view
device location information for TM80 telemetry monitors. For details on how to import AP
information, see
BeneVision Central Monitoring System Operator’s Manual
. To view
device location, follow this procedure:
1.
Select the
symbol at the top of the ViewBed screen. The
Device Location
menu
is displayed.
2.
View the AP switch time in the
Time
column and the current location of devices in
the
Location
column.
3.
If you wish to locate aTM80 telemetry monitor, select the
Find Device
button. The
TM80 will generate a continuous audible tone until it is acknowledged at the TM80.
WARNING
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...