WARNING
•
If you switch off all arrhythmia alarms, the monitor will not alarm for any
arrhythmia event. This may result in a hazard to the patient. Always keep the
patient under close surveillance.
NOTE
•
The priority of lethal arrhythmia alarms is always high.
•
When the telemetry monitor is connected to the CMS, any changes made either
at the telemetry monitor or the CMS will be communicated to the other side.
•
When lethal arrhythmia alarms are set to Disable at the CMS, the
symbol is
displayed on the lethal arrhythmia alarms at the telemetry monitor.
•
If any of the lethal arrhythmia alarms is switched off, the ECG waveform area at
the telemetry monitor displays the “Lethal Arrhys Off” message. The ECG
waveform area on the screen of the CMS displays Asystole, V-Fib/V-Tach, or V-
Tach Off depending on the actual lethal arrhythmia alarm.
Irr Rhythm
Configures arrhythmia
alarms.
On,
Off
AF/Irr Rhy End Time: 0 min, 1 min,
2 min
, 3
min, 4 min, 5 min, 10 min, 15 min, 30 min
Alarm priority:
Prompt
, Low
A-Fib
On,
Off
AF/Irr Rhy End Time: 0 min, 1 min,
2 min
, 3
min, 4 min, 5 min, 10 min, 15 min, 30 min
Alarm priority:
Prompt
, Low, Med, High
PVCs/min
On,
Off
PVCs /min: 1 to 100; the default is
10
.
Alarm priority: Prompt, Low,
Med
, High
Pauses/min
On,
Off
Pauses/min: 1 to 15; the default is
8
.
Alarm priority: Prompt, Low,
Med
, High
Options
Description
Settings*
*
The factory default settings are in bold.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...