BeneVision TM80 Telemetry Monitor Operator’s Manual
15 - 11
SpO
2
SpO
2
Sensor Off
Low*
B
The SpO
2
sensor
has become
detached from
the patient or
the module.
There is a fault
with the SpO
2
sensor.
An unspecified
SpO
2
sensor has
been used.
Check the sensor
application site
and the sensor
type, and make
sure the sensor is
not damaged.
Reconnect the
sensor or use a
new sensor.
SpO
2
Sensor
Fault
Low
C
SpO
2
No
Sensor
Low
B
SpO
2
Module
Error
High
A
SpO
2
Excess
Light
Low
A
There is too
much light on
the SpO
2
sensor.
Move the sensor
to a place with
lower level of
ambient light or
cover the sensor
to minimize the
ambient light.
SpO
2
No
Pulse
Low
A
SpO
2
sensor
failed to obtain
pulse signal.
Move the sensor
to a site with
better perfusion.
SpO
2
SpO
2
Unplugge
d
Low
C
SpO
2
module
connector is
disconnected
from the
telemetry
monitor.
Reconnect the
SpO
2
module to
the telemetry
monitor.
SpO
2
Sensor
Incompati
ble
Low
C
Incompatible or
an unspecified
SpO
2
sensor is
used.
Use specified
SpO
2
sensors.
Measurement
Alarm
message
Default
Alarm
priority
Alarm
Indication
Possible cause
Solution
*
The asterisk (*) means the alarm priority is configurable.
**
XX is the ECG lead name.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...