BeneVision TM80 Telemetry Monitor Operator’s Manual
10 - 7
10.5.4 Configuring NIBP Alarm Settings
To configure NIBP alarm settings, follow this procedure:
1.
In the
Alarms
section of the
NIBP
menu, tap
NIBP Alarm Setup
.
2.
Select the options described in the following table.
Options
Description
Settings*
Mode
Configures NIBP measurement
mode.
Manual, Auto, Seq., ABPM
Initial Pressure
Configures the initial cuff pressure.
For adult:
80 mmHg to 280 mmHg; the
default is
160
.
For pediatric:
80 mmHg to 210 mmHg; the
default is
140
.
Color
Configures the color of NIBP
numeric data.
16 colors
The default color is white.
*
The factory default settings are in bold.
Options
Description
Settings*
SYS
Configures whether to trigger
the NIBP systolic pressure
alarm.
On
, Off
Alarm high limit:
• For adult:
Range: (low limit + 5) to 289
Default value:
160
• For pediatric:
Range: (low limit + 5) to 239
Default:
120
Alarm low limit:
• For adult:
Range: 26 to (high limit - 5)
Default value:
90
• For pediatric:
Range: 26 to (high limit - 5)
Default:
70
Alarm priority
:
Med,
High
*
The factory default settings are in bold.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...