BeneVision TM80 Telemetry Monitor Operator’s Manual
NOTE
•
When the telemetry monitor is connected to the CMS, the patient category at
the CMS is updated if the patient category is changed at the telemetry monitor.
Refer to
BeneVision Central Monitoring System Operator’s Manual
for details.
5.3.3 Changing the Paced Status
It is important to correctly set the patient’s paced status before patient monitoring.
To change paced status, follow this procedure:
1.
In the main menu, tap
Patient Info
.
2.
Check if
Paced
is appropriate for the patient.
3.
If the paced status setting is not correct, tap
Paced
and select the correct option.
◆
When
Paced
is set to
Yes
at the telemetry monitor, if pace pulses are detected,
the
symbol displays in the waveform area of the Central Monitoring
System’s screen, and the pace pulse marks will display on the ECG waveform
both at the telemetry monitor and Central Station.
◆
When
Paced
is set to
No
or not specified at the telemetry monitor, the
symbol displays in the waveform area of the Central Monitoring System’s
screen.
WARNING
•
For paced patients, you must set Paced to Yes. If it is incorrectly set to No, the
Central Station could mistake a pace pulse for a QRS and fail to alarm when the
ECG signal is too weak.
•
On ventricular paced patients, episodes of ventricular tachycardia may not
always be detected.
•
Do not rely entirely upon the system’s automated arrhythmia detection
algorithm. Keep paced patients under close surveillance.
•
For non-paced patients, you must set Paced to No. The pacer pulses may be
counted as QRS complexes, hence leading to wrong HR readings or failure to
diagnose certain arrhythmia symptoms. Be sure to keep a close eye on patient’s
with pacemaker devices.
•
The radio frequency energy of the telemetry monitor or other radio frequency
sources, when used in close proximity to a pacemaker, may interfere with
pacemaker performance. Internal pacemakers are less vulnerable than external
pacemakers due to the shielding effects of the body. However, caution should
be exercised when monitoring any paced patient.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...