PN B10093AE
3-5
Troubleshooting
Accessing the Rotor in Case of Power Failure
3
To access the door-locking mechanism, you must remove the instrument front panel. Two latches
secure the front panel in place; these latches are accessible through two holes at the upper right
and left of the panel (see
).
Figure 3.1
Emergency Door Release Latch Access
1
Turn the power switch to off (
O
) and unplug the power cord from the main source by removing
the Mains (power) plug from the outlet receptacle.
•
The Mains (power) plug is the disconnect device and must remain easily accessible.
—
Position the centrifuge so that it is easy to remove the Mains (power) plug from the
outlet receptacle.
2
Insert a 4-mm or
5
/32-in. Allen wrench straight through one of the holes (either one) and turn
the wrench to the left (counterclockwise) until the latch disengages.
3
Repeat step 2, inserting the Allen wrench through the other hole.
•
After the second latch disengages, the front panel will fall forward about an inch from
the top.
•
The bottom of the front panel is held in place by three grooved brackets, attached to the
front panel, that seat over a metal retaining strip on the centrifuge bottom panel.
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......