PN B10093AE
4-2
Care and Maintenance
Maintenance
3
Wipe condensation out of the chamber between runs with a sponge or clean cloth to prevent
chamber icing.
4
If chamber icing occurs, defrost the system and wipe moisture out of the chamber before use.
a.
To defrost the system, set the temperature to 30°C for 20 minutes.
(These are suggested settings that may be adjusted as appropriate for your laboratory
conditions.)
NOTE
Before using any cleaning or decontamination methods except those recommended by the
manufacturer, users should check with the manufacturer that the proposed method will not damage
the equipment.
Cleaning
1
Clean the centrifuge frequently.
a.
Always clean up spills when they occur to prevent corrosives or contaminants from drying
on component surfaces.
2
To prevent accumulations of sample, dust, and/or glass particles from broken sample tubes,
keep the chamber clean and dry by frequent wiping with a cloth or paper towel.
a.
For thorough cleaning, wash the chamber using a mild detergent such as Solution 555.
b.
Dilute the detergent with water (10 parts water to 1 part detergent).
c.
Rinse thoroughly and dry completely.
3
Clean the centrifuge exterior surfaces by wiping with a cloth dampened with Solution 555.
Dilute the detergent with water (10 parts water to 1 part detergent).
Do not use acetone.
4
Clean the drive hub regularly using Solution 555 and a soft brush.
a.
Dilute the detergent with water (10 parts water to 1 part detergent).
b.
Rinse thoroughly and dry completely.
Tube Breakage
1
If a glass tube breaks, and all the glass is not contained in the bucket or rotor, be sure to
thoroughly clean the chamber.
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......