PN B10093AE
3-2
Troubleshooting
User Messages
Table 3.1
Diagnostic Message Chart
Diagnostic Message Problem
Result
Recommendation
P1/
Power failure, run
continued
Momentary power failure:
rotor does not come to a
complete stop
Run continues when
power resumes
Press
CE
to clear message.
P2
/
Power failure, run
restarted
Power failure: rotor comes
to a complete stop
Run will restart
automatically when
power resumes
Press
CE
to clear message.
L2
/
Door not latched,
reclose door
L5
/
Door not latched, call
Beckman service
Door not closed firmly
enough, or debris in latch
area may be preventing
latches from closing
securely
L2 error message
appears. If you reclose
the door repeatedly and
the problem continues,
the message will change
to say “
L5 Door not
latched, Call Beckman
Service
”
a
Reclose the door firmly, placing your
hand over the drawing of the hand on
the door label. (As you close the door
your hand should be half-way between
the two latches, which is approximately
45° to the right of center front.)
Gently clean the latch areas with a dry,
lintless cloth. Be careful not to damage
sensitive electronics in the area. If L2
continues to appear after cleaning, call
Beckman Coulter Field Service.
a
L3
/
Door problem, refer to
manual
L9
/
Door problem, refer to
manual
Foot pedal was depressed
before end of run
Latches are not
operating properly
Press
CE
and try to restart the run.
If this does not work, call Beckman
Coulter Field Service.
a
L4
/
Door problem, refer to
manual
Door stays locked after
rotor stops spinning
Sample cannot be
retrieved
Restart the centrifuge and perform a
brief run. If door will still not open after
decel, call Beckman Coulter Field
Service.
a
C2
/
Rotor over temp, run
stopped
Rotor temperature
exceeds maximum
temperature setting
Run shuts down with
maximum brake
Call Beckman Coulter Field Service.
a
C3
/
No temp control, refer
to manual
Refrigeration system
malfunction
Run shuts down with
maximum brake
•
Check the air filter and replace if
).
•
Call Beckman Coulter Field Service.
a
D1
and
D9
Erroneous signal from
tachometer board to
display board or drive
system problem. Door
lockout time may not be
sufficient to prevent access
to spinning rotor.
Erroneous display of
0 RPM immediately after
failure, rotor still
spinning.
Call Beckman Coulter Field Service.
a
Wait for 4 hours before trying to open
the door.
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......