PN B10093AE
2-12
Operation
Manual Operation
2
Press the appropriate soft key.
•
If you choose
MAX
or
SLOW
, your selection appears in the
Accel:
field.
•
If you choose
TIME
, enter an acceleration time from 1 to 10 minutes and press
A/D
again or
accept the 10-minute default setting.
Deceleration
1
Press
A/D
twice (if the cursor is in the
Accel:
field, press
A/D
once).
The
Decel:
field flashes.
MAX
,
SLOW,
TIME,
and
OFF
soft keys appear.
2
Press the appropriate soft key.
•
If you choose
MAX
,
SLOW,
or
OFF
, your selection appears in the
Decel:
field.
•
If you choose
TIME
, enter a deceleration time from 1 to 10 minutes and press
A/D
again or
accept the 10-minute default setting.
Starting a Run
1
Press
ENTER
, then
START
.
•
(You must press
START
within 5 seconds of pressing
ENTER
.)
•
The green light on the
START
key flashes and the rotor begins to spin.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Rotor : JA-30.50
Speed : 26 000 RPM = 81 800 x G
Time : 00:25
Temp : 20°C Max Temp : 24°C Shutdown? No
MAX
SLOW
TIME
Accel : MAX Decel :
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Rotor : JA-30.50
Speed : 26 000 RPM = 81 800 x G
Time : 00:20
Temp : 20°C Max Temp : 24°C Shutdown? No
MAX
SLOW
TIME
OFF
Accel : MAX Decel : 5
Enter 1 to 10 minutes from 500 to 0 rpm
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......