PN B10093AE
2-2
Operation
Summary of Avanti J-26S XPI Run Procedures
2
Depress the foot pedal to open the chamber door.
3
Install the rotor according to instructions in the applicable rotor manual, then firmly close the
chamber door.
4
Press
ROTOR,
press the soft key to select rotor type, press the soft key to select the rotor name.
5
Press
SPEED,
press the soft key to select
RPM
or
RCF
mode, then use the keypad to enter the run
speed (0 to 26,000 RPM).
6
Press
TIME
press the soft key to select a time mode (
HH:MM
,
Hold
or
2
t
), then use the keypad
to enter the run time or
2
t
value.
(No entry is made in
Hold
mode.)
7
Press
TEMP
then use the keypad to enter the required run temperature.
8
Press
A/D
and press the
MAX
,
SLOW,
or
TIME
soft key. Or, skip this step and accept the default
acceleration rate (
MAX
).
9
Press
A/D
and press the
MAX
,
SLOW,
TIME
, or
OFF
soft key.
a.
(Or, skip this step and accept the default deceleration rate (
MAX
).
10
Check that all parameters are correct and that the door is closed.
a.
Press
ENTER,
then press
START.
11
Wait for the set time to count down to zero, or end the run by pressing
STOP
.
12
When the rotor stops, depress the foot pedal to open the chamber door.
Programmed Run
1
Press the
POWER
switch to on (
I
).
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......