Description
Specifications
Keypad
The keypad is used to enter numerical run parameters and to enter or recall a program number. In
addition to keys 0 through 9, the keypad includes
.
(plus/minus/decimal) and
CE
(clear entry) keys.
The
.
key is used to enter a decimal point when typing in a rotor number (for example, 25.50) and
enter a temperature setting below 0°C (for example, –2°C).
The
CE
(clear entry) key clears any parameter in the active field. Pressing
CE
will also clear some
diagnostic messages. If a diagnostic message does not clear when
CE
System Keys
Specifications
Only values with tolerances or limits are guaranteed data. Values without tolerances are informative data,
without guarantee.
A/D
Pressed to enter acceleration and deceleration rates that will maintain optimum separation
while protecting samples. When
A/D
is pressed, the
Accel:
field flashes, prompting you to
enter an acceleration setting. Pressing
A/D
once more moves the cursor to the
Decel:
field.
Available acceleration rates are maximum, slow, and timed (1 to 10 minutes from 0 to
500 RPM). Available deceleration rates are maximum, slow, timed (1 to 10 minutes from
500 to 0 RPM), and no brake (off).
OPT
Pressed to put the centrifuge into program mode, where you can enter run parameters into
memory for later use. Up to 30-step programs can be stored.
START
Pressing
ENTER
, then
START
begins the run.
START
must be pressed within 5 seconds
of pressing
ENTER
.) This key combination can also be used to abort a deceleration
process and restart the centrifuge.
STOP
Pressed to end a run according to the deceleration setting. The red light on the
STOP
key flashes while the rotor decelerates. An audible tone sounds when the rotor comes
to a complete stop. Deceleration can be terminated and the run restarted by pressing
ENTER
, then
START
.
During a zonal run, pressing
STOP
once will decelerate the rotor to 2000 RPM. Pressing
STOP
again will decelerate the rotor to 0 RPM.
ENTER
Pressed to save parameter inputs to system memory, and to start the centrifuge
(
ENTER
, then
START
must be pressed to start the centrifuge).
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......