PN B10093AE
2-8
Operation
Manual Operation
Entered time appears in the
Time:
field.
3
Press the next function key or press
ENTER
, then
START
.
•
The run begins, and the time remaining begins counting down to 0 in the
TIME
display.
•
If you enter a number over 59 minutes, the centrifuge automatically recalculates the time
in hours and minutes after
ENTER
or any function key is pressed.
•
Deceleration begins when the
TIME
display has counted down to 0.
•
During deceleration, “
End
” flashes in the
TIME
display and the red indicator light on the
STOP
key flashes.
Entering a
2
t
Run
1
Press
TIME
.
HH:MM
,
W2T,
and
HOLD
soft keys appear.
2
Press the
W2T
soft key to select the
2
t
mode.
The
Time:
field flashes.
3
Use the keypad to enter an
2
t
value.
•
The entry appears in the
Time:
field.
•
The decimal point and
E
appear automatically as you type.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Rotor : JA-30.50
Speed : 26 000 RPM = x G
Time : 00:25
Temp : 25°C Max Temp : 29°C
Accel : MAX Decel : MAX
HH:MM
W2T
HOLD
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......