PN B10093AE
2-4
Operation
Manual Operation
CAUTION
Do not drop the rotor onto the drive hub. The drive shaft can be bent if the rotor
is forced sideways or dropped onto the spindle hub. Install the rotor by centering
it over the hub and carefully lowering it straight down.
4
Securely attach the rotor lid knob, or tie-down knob in rotors without lids, to the drive shaft by
turning it to the right (clockwise).
NOTE
If the knob turns loosely and you do not feel threads engaging, the rotor drive hole pins may not
be properly seated on the centrifuge hub. Lift the rotor up, rotate it slightly, and lower it onto the hub
again. Tighten the knob.
5
Close the chamber door.
CAUTION
If you leave the rotor in the centrifuge between runs, make sure the rotor is seated
on the drive hub and the tie-down knob is tight before each run.
Manual Operation
•
When a function key (
ROTOR
,
SPEED
,
TIME
,
TEMP,
A/D
, or
OPT
) is pressed, the corresponding
setup screen field blinks to indicate that a parameter can be entered or changed. The field
continues to blink until
ENTER
or another function key is pressed.
•
To change an entry before you’ve pressed
ENTER
press
CE
and enter a different value. To change
an entry after you’ve pressed
ENTER
, press the function key again.
•
If an invalid setting is entered, the display blinks and the valid range for that parameter is
displayed in the message line. A valid parameter must be entered and
ENTER
must be pressed
before you can enter another parameter.
Selecting a Rotor
1
Press
ROTOR
.
The
Rotor:
field flashes, and rotor type abbreviation soft keys (
JA:
fixed angle,
JS:
swinging bucket,
JLA:
lightweight fixed angle;
JSP:
special;
JV:
vertical tube;
JE:
elutriator;
JCF-Z:
continuous flow/zonal) appear.
2
Press the appropriate soft key to select the rotor type.
A list of rotor codes appears—fixed angle rotor codes are displayed below.
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......