PN B10093AE
2-13
Operation
Programmed Operation
2
Changing Parameters During a Run
While a run is in progress, run parameters (speed, time, temperature, and acceleration and
deceleration settings) can be changed without stopping the run by entering the new setting and
pressing
ENTER
.
Ending a Run
To end a run in progress, press
STOP
. The red
STOP
indicator light flashes until the rotor comes to
a complete stop.
In timed runs, deceleration begins automatically when the
TIME
display counts down to zero. In
2
t
runs, deceleration begins when the set
2
t
is achieved.
NOTE
If a diagnostic condition causes the run to stop, a diagnostic message appears and the red
STOP
indicator light flashes until the rotor comes to a complete stop. See
for diagnostic code
explanations and troubleshooting procedures.
Programmed Operation
The centrifuge internal memory can store up to 30 programs, each containing up to two steps (each
step is a set of run parameters). Programs are retained in memory when the power is turned off.
Creating a New Program
1
Press
OPT
.
The
Prog
(ram)
Recall:
field flashes, and the message line says “
Enter program number (1–30)
.”
2
Press
OPT
again to move the cursor to the
setup:
field.
a.
Use the keypad to enter a program number in this field, then press
ENTER
.
•
A two-step program is created and is assigned the number you entered.
•
If you enter a number that is already assigned to a program, you can either overwrite
the program or select another program number.
•
A message prompts you to enter parameters for step 1.
Rotor : JA-30.50
Speed : 0 RPM = x G
Time : 00:00
Temp : 25°C Max Temp : 29°C Shutdown? No
STEP1
STEP2
SAVE
EXIT
Accel : MAX Decel : MAX
Prog Recall: Setup: 1 Step #1
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Enter parameters for the selected step
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......